LQ-HXP 120
LEJ

External Light Source for Fluorescence Excitation Microscopy
This compact device is the ideal illumination source for fluorescence excitation in microscopy. Coupling is preferably achieved via liquid light guides to reduce heat at the microscope and to allow for remote shuttering of the lamp, reducing potential vibrations during time-lapse imaging. The average service life of a lamp is 2,000 hours. The optimized internal air flow cools the lamp and enables quiet operation. A resettable operating hour meter shows lamp operating hours. The brightness of the lamp output can be varied in 5 steps (0-100%) by means of a mechanical dimmer unit. An ultra-fast shutter opens or closes after activation with approx. 6 ms. The shutter is actuated either by means of an appliance switch, or externally by means of a remote control input or CAN-BUS connection. This compact light source is available as a VIS or UV version in addition to a version that permits the coupling of multi-pin liquid light guides.
Electrical Connection |
Mains voltage |
100 up to 240 VAC ± 10 %, 50 up to 60 Hz |
Power input |
max. 210 VA |
Configuration |
Mercury short-arc lamp |
HXP R 120W/45C VIS with focussed reflector (Osram) |
Filter / heat protection filter installed in filter holder |
UV-IR Cut - Filter |
Shutter |
electrical control via interface, max. frequency 40 Hz, switching time approx. 6 ms |
Dimming |
mechanically, 5 steps 0 ... 100% |
Operating Hour Meter |
display of lamp operation hours, resettable |
Interface: Shutter Activation |
remote control connection for remote signal and foot switch, CAN-BUS-connector on request |
Desktop Device |
Dimensions (HxWxD) |
210 x 130 x 290 in mm |
Weight |
ca. 4,7 kg |
Safety |
as per EN 61010-1 |
Equipment Class |
I |
Protection Class |
IP 20 (EN 60529) |
Safety Circuit |
switches unit supply off when device lid is opened |
Excess Temperature Protection |
switches lamp voltage off in the event of excess temperature, switches on again after cooling |
Glare Shield |
light output remains automatically closed as long as no light guide is connected |
EMC |
Interference |
EN 61326 (class B) |
Immunity |
EN 61326 appendix A |
Voltage Fluctuations, Flicker |
EN 61000-3-3 |
Harmonic Current Emissions |
EN 61000-3-2 |
Environmental Conditions |
Ambiente Temperature |
0 °C to 40 °C |
Storage Temperature |
-20 °C to + 85 °C |
Conformity |
|
CE |
|
RoHS |
Approvals |
|
UL/CSA |
Lightning |
Replacement Bulbs |
HXP R 120W/45C VIS |
Adapter |
UV-Lighting Adapter |
Leica Motik Nikon Olympus Zeiss for other microscopes on request |
Liquid-Light-Guide LUMATEC Series 300 |
1-pin light input: Storz long, light output: Standard |
Ø 3 mm x 2.000 mm Ø 5 mm x 1.500 mm other lengths on request |
2-pin light input: LUMATEC D, light output: 2 x Standard |
2 x Ø 3 mm x 1.500 mm other lengths on request |
Filter |
Filter / Heat Protection Filter |
IR reflection filter |
Other Filters |
on request |
Filter Holder |
on request |
Remote Control |
Shutter Remote Control |
hand / foot-actuated switch |
Remote-Cable |
on request |
|
|